Baxter 6201 Infusion Pump

Baxter

$675.00
(No reviews yet) Write a Review

No Interest if paid in 6 months on purchases of $99+

SKU:
6201
Condition:
Refurbished
Weight:
11.80 LBS
Shipping:
Calculated at Checkout
Adding to cart… The item has been added

Operator's Manual PDF Download

The Baxter FLO-GARD 6201 Single-Channel Volumetric Infusion Pump can deliver a wide variety of fluids over a broad range of infusion rates.
The FLO-GARD 6201 Volumetric Infusion Pump and the FLO-GARD 6301 Dual-Channel Volumetric Infusion Pump are industry standard in standard set infusion pump technology. 

Enhanced Safety

  • Flow Check Occlusion Alarm offers an in-line resistance display of incremental back pressure.
  • Slide Clamp option offers an additional step to protect against inadvertent gravity free flow.
  • Flow Rate Calculation is automatic after volume and time are selected.
  • Front Panel Lock-out protects against tampering. 

Greater Convenience and Versatility

  • Programmed Delivery Profile allows up to 10 steps, individualizing control of infusion ramping and tapering.
  • Incremental Flow Rate can deliver 1-99.9 mL/hr in increments of 0.1 mL/hr.
  • Automatic Restart once occlusion clears.
  • Automatic Piggybacking of secondary medications. 

Cost-Efficiency

  • Uses Baxter standard IV administration sets that eliminates the need for costly disposable sets and reduces the potential for waste.
Flow Rate Range: 1-99.9 mL/hr in 0.1 mL/hr increments and 1-1,999 mL/hr in 1 mL/hr increments
VTBI Range: 1-9,999 mL
Alarm Log: Last 10 alarms
Dimensions: 20 cm W x 13 cm D x 29 cm H (8 inches x 5.1 inches by 11.4 inches)
Weight: 5.3 kg (11.6 lbs)
Battery: 12 volt, 2.0 ampere, sealed lead acid
Battery Recharger: 80% charged within 8 hours
Battery Life: 6 hours at 1,400 mL/hr
Power cord: Biomed replaceable 2.9 m (9 feet)
Leakage Current: Less than 50 µA
Fuse: ½ ampere, 250 volt, slow-blow
Communications Port: RS232 serial port
Power Requirements: 115 VAC, 60 Hz, 30W
1 Year Warranty